Versa Controls
AX60+




Click image to zoom

Specifications:

System Warm-up time: 40 seconds Operating temperature range: -5 °C to +50 °C, +23 °F to +122 °F Approvals: CE, EN50270, IEC 61010 (UL), DIN 6653 (TUV)
System Warm-up time: 40 seconds Operating temperature range: -5 °C to +50 °C, +23 °F to +122 °F Approvals: CE, EN50270, IEC 61010 (UL), DIN 6653 (TUV)
Inlet pressure range 1 to 3 bar gauge (14.5 to 42.5 psi)
Calibration gas flow rate 0.5 to 1.0 litres/min (approx)
Sensor accuracy better than ± 1 ppm at constant temperature and pressure*
Response time < 90 seconds
Operating temperature 0 to 40 °C (32 to 104 °F)
Temperature effect 0.04% of reading/ °C
Atmospheric pressure range 800 mbar to 1200 mbar
Warm up time 2 minutes
Weight 400 g / 0.8 lbs
Dimensions 175 x 130 x 70 mm (l x w x d), 6.8 x 5.11 x 2.75 inches (l x w x d)
IP rating IP65
Sensor type electrochemical cell
Sensor life (expected) 2 years at standard temperature and pressure
Sensor warranty 9 months graded
Electronics warranty 1 year
Display 4 digit LCD
Alarms 2 x alarm visual indicators, 1 x system fault indicator, 1 x status indicator, common audible alarm
Relays one or two optional alarm relays with changeover contacts assigned to alarm 1, alarm 2 or system fault, contact rating 230 V AC or 30 V DC at up to 2 A, contacts are non-latching fail-safe
Outputs 2 wire 4-20 mA (max load 150 O)
Power supply options Option selected at time of manufacture/order, 210/250 V AC, 110/120 V AC, 9-24 V DC
Product Information:

The Ax60+ is a wall-mountable gas safety device for monitoring oxygen and carbon dioxide. Based on the popular Ax60 CO2 detector, the new Ax60+ offers the additional functionality of a modular O2 sensor which provides an early warning of both oxygen depletion and oxygen enrichment. The CO2 and O2 sensors are interchangeable and can be fully integrated as part of a multi-point, multi-gas detection and alarm system. The CO2 sensor is set by default to trigger a low-level alarm at 1.5% CO2, an evacuation alarm at 3% CO2 and a time-weighted average alarm of 0.5% CO2 measured over eight hours. The O2 sensor is set by default to trigger low-going alarms at both 19.5% and at 18% and a high-going alarm at 23%. The alarm set points can be changed by the user. Warnings are announced by high-visibility strobe lights and high-volume sounders.




Distributors of



Contact Us


Address :

Survey No. 143/1/D,
Milkat No. 2450/1,
Village Vadgaon - 412106,
Tal.: Maval, Dist.: Pune,
Maharashtra, India

Office:

+91-7722020796
info@versacontrols.com


Sales:

+91-7447797456
sales@versacontrols.com





© 2019 Versa Controls. All Rights Reserved.
Powered by aplap software